  agilent U2701A and u2702a usb modular oscilloscope data sheet
2 features 100 mhz and 200 mhz bandwidths up to 1gsa/s maximum sample rate 32 mpts of waveform memory compact and portable size 117.00 mm x 180.00 mm x 41.00 mm (with rubber bumpers) advanced triggering, including edge, pulse width, and tv compatibility with hi-speed usb 2.0, usbtmc 488.2 standards four math functions, including ffts standard dual-play standalone and modular capability compatibility with wide range of application development environments highest performance, lowest cost for today and tomorrow agilents U2701A and u2702a usb modular oscilloscopes combine a set of essential features that are ideal for analyzing designs in an affordable way. the U2701A and u2702a come in two bandwidths: 100 mhz and 200 mhz respectively. these devices are uniquely designed to accommodate your needs for ? exibility with the dual- play function. dual-play functionality allows you to use the oscilloscope as a standalone or to scale up the test system in a cardcage with additional scopes or with agilents other usb modular product offerings, thus providing a complete solution for system development. the U2701A and u2702a give you the debugging power you need. each modular oscilloscopes comes standard with features such as advanced triggering, automatic measurements, math functions including ffts, and much more. the user-friendly agilent modular instrument measurement manager software bundled with the U2701A and u2702a offers a simple interface for quick setup, con? guration, and measurement control. figure 1 the U2701A and u2702a usb modullar oscilloscopes
3 why do you need deep memory and a high sampling rate? to see more time when you are able to store more samples that you have obtained in the memory, you can view the signal at a longer time. this will be the best way to understand the use of deep memory. a longer capture time gives you a better visibility into cause-effect relationships in your designs, which signi? cantly simplify your root-cause debugging. it also allows you to capture start-up events in a single acquisition. the need to stitch together multiple acquisitions or set precise triggering conditions are no longer necessary. you can spend less time ?nding events, and more time analyzing them. to see even more details the relationship between memory depth and acquisition rate is not as obvious. all scopes have a banner maximum sample rate speci? cation, but many can only sustain these rates at a few time base settings. higher sampling rate by offering sampling rate more than twice the acquired signal bandwidth, aliasing can be prevented. with more sampling data captured, higher accuracy of your test and analysis results can be achieved. ease of use the U2701A and u2702a usb modular oscilloscopes are equipped with hi- speed usb 2.0 interface for easy setup and plug-and-play. hence, this ease-of- use makes the oscilloscopes ideal for the education, design validation, and manufacturing environment. compatible with a wide range of application development environments the agilent U2701A and u2702a usb modular oscilloscopes are compatible with a wide range of application development environments. this minimizes the time that r&d and manufacturing engineers need to use the devices in different software environments. listed below are the popular development environments and tools with which the usb modular oscilloscope is compatible: ? agilent vee and agilent t&m toolkit ? microsoft visual studio.net, c/c++ and visual basic 6 ? labview ? microsoft .net framework figure 3 the accessories offered for the U2701A and u2702a usb modular oscilloscopes. figure 2 the dual-play capability allows U2701A and u2702a usb modular oscilloscopes to be used as standalone units or ? tted into a cardcage.
4 features you need the U2701A and u2702a include the following standard features that you need to do your job done quickly and easily: hi-speed usb interface the U2701A and u2702a connect to the computer through hi-speed usb 2.0 connectivity. autoscale autoscale lets you display any active signals, automatically setting the vertical, horizontal, and trigger controls for the best signal display within the shortest time. advanced triggering edge, pulse width, and tv are the triggering modes included to help you isolate the signals you want to see. large memory with memory depth up to 32 mpts, you can capture even more data. larger memory allows you to capture data over a longer time frame. fast fourier transfer (fft) and waveform math the U2701A and u2702a offer analysis functions such as addition, subtraction, multiplication, division, and fast fourier transform (fft). fft allows you to manipulate the waveform using ? ve types of windows such as hanning, hamming, blackman-harris, flattop, and rectangular. figure 4 the U2701A and u2702a connect to the computer or laptop with a usb cable, enabling fast data transfer. high sampling rate sampling rate up to 500 msa/s/ch enables more details of the signal to be seen and analyzed. when two channels are interleaved, the sampling rate can be up to 1 gsa/s. this fast- sampling capability allows you to perform intermittent detection easily. pulse triggering pulse triggering allows you to trigger on pulse events. portability the U2701A and u2702as compact size makes them portable and easy to be carried to your working ? eld. one-year warranty every U2701A and u2702a comes with one year warranty.
5 product characteristics and general speci? cations remote interface hi-speed usb 2.0 usbtmc class device power consumption +12 v dc, 2 a installation category iii operating environment operating temperature: 0 c to 50 c storage temperature: C20 c to 70 c operating humidity: 20 ~ 85% r.h. non-condensing storage humidity: 5 ~ 90% r.h. non-condensing altitude: up to 2,000 m (operating and non-operating) pollution degree: 2 storage compliance C20 c to 70 c safety compliance certi? ed with: iec 61010-1:2001/ en61010-1:2001 (2nd edition) canada : can/csa-c22.2 no. 61010-1-04 usa: ansi/ul 61010-1:2004 emc compliance iec 61326-2002/ en 61326:1997+a1:1998+a2:2001+a3:2003 canada : ices-001:2004 australia/new zealand: as/nzs cispr11:2004 shock and vibration tested to iec/en 60068-2 io connector bnc connector dimension (w x d x h) 117.00 mm x 180.00 mm x 41.00 mm (with rubber bumpers) 105.00 mm x 175.00 mm x 25.00 mm (without rubber bumpers) weight 534 g (with rubber bumpers) 482 g (without rubber bumpers) warranty one year system requirements processor 1.6 ghz pentium iv or higher operating system one of the following microsoft? windows? versions: windows xp professional or home edition (service pack 1 or later) windows 2000 professional (service pack 4 or later) browser microsoft internet explorer 5.01 or higher available ram 512 mb or higher recommended hard disk space 1 gb prerequisites agilent io libraries suite 14.2 [1] or higher agilent t&m toolkit 2.1 runtime version [2] microsoft .net framework version 1.1 and 2.0 [2] [1] available in agilent automation-ready cd [2] bundled with agilent measurement manager software application installer
6 standard shipped items ? 12 v, 2 a ac/dc power adapter ? power cord ? usb standard a to mini-b interface cable ? 10:1 passive probe 150 mhz 1.2m, n2862a (only applicable for U2701A) ? 10:1 passive probe 300 mhz 1.2m, n2863a (only applicable for u2702a) ? l-mount kit ( used with modular instrument chassis) ? agilent automation-ready cd (contains the agilent io libraries suite) ? agilent usb modular products quick start guide ? agilent usb modular products reference cd-rom ? agilent usb modular products quick reference card ? certi? cate of calibration optional accessories ? bnc cable, u2921a-100 ? usb secure cable, u2921a-101 ? 1:1 passive probe 20 mhz, 1.5 m, 10070c (order no.: U2701A-200) ? 10:1 passive probe 150 mhz 1.2m, n2862a (only applicable for U2701A) (order no.: u2701-201) ? 10:1 passive probe 300 mhz 1.2m, n2863a (only applicable for u2702a) (order no.: u2702-201) product outlook and dimensions front view rear view top view 41.00 mm 180.00 mm 15.00 mm 117.00 mm
7 performance speci? cations [1] 1 all speci? cations are warranted. speci? cations are valid after a 30-minute warm-up period and within 10oc of last calibration temperature. vertical system: oscilloscope channels bandwidth (C3 db) U2701A: dc to 100 mhz u2702a: dc to 200 mhz scope channel triggering trigger sensitivity <10 mv/div: greater of 1 div or 5mv; 10 mv/div: 0.6 div
8 performance characteristics [1] 1 all characteristics are typical performance values and are not warranted. characteristics are valid after a 30-minute warm-up period and within 10oc of last calibration temperature. 2 under standalone use, users are only allowed to measure up to cat1 30 vrms. for high-voltage measurement up to cat1 300 vrms, users must install the l-mount kit on the U2701A/u2702a before plugging into the instrument chassis. ensure that the l-mount kit installed on your modular oscilloscope is screwed to the instrument chassis to ensure chassis grounding. i t is required to use the provided 10:1 probes (n2862a/n2863a) for high-voltage measurements acquisition: oscilloscope channels real time sample rate 2 channels interleaved each channel 1 gsa/s 500 msa/s standard memory depth 2 channels interleaved each channel normal single-shot 32 mpts 64 mpts 16 mpts 32 mpts vertical resolution 8 bits peak detection yes averaging any number from 1 to 999 filter sin(x)/x interpolation for time base 1 ns to 100 ns sweep modes auto, normal, single vertical system: oscilloscope channels scope channels U2701A/u2702a: ch 1 and 2 simultaneous acquisition ac coupled U2701A: 3.5 hz to 100 mhz u2702a: 3.5 hz to 200 mhz calculated rise time (= 0.35/bandwidth) U2701A: 3.5 ns u2702a: 1.75 ns single-shot bandwidth U2701A: 100 mhz u2702a: 200 mhz range 2 mv/div to 5 v/div (1 m?) maximum input [2] cat i 30 vrms, 42 vpk offset range 4 div example: 8 mv on 2 mv/div; 20 v on 5 v/div dynamic range 4 div input impedance 1 m ? : _ 16 pf coupling ac, dc, ground bw limit _ 25 mhz standard probes 10:1 passive probe 150 mhz 1.2 m 10:1 passive probe 300 mhz 1.2 m esd tolerance 2 kv noise peak-to-peak 3 mvpp dc vertical offset accuracy 200 mv/div: 0.1 div 2.0 mv 0.5% offset value; > 200 mv/div: 0.1 div 2.0 mv 1.5% offset value dc vertical gain accuracy 4.0% of full scale
9 performance characteristics (continued) vertical system: oscilloscope channels (continued) single-cursor accuracy {dc vertical gain accuracy + dc vertical offset accuracy + 0.2% full scale (~1/2 lsb)} example: for 50 mv signal, scope set to 10 mv/div (80 mv full scale), 5 mv offset, accuracy = {4.0% (80 mv) + 0.1(10 mv) + 2.0 mv + 0.5% (5 mv) + 0.2% (80 mv)} = 6.385 mv dual-cursor accuracy {dc vertical gain accuracy + 0.4% full scale (~1 lsb)} example: for 50 mv signal, scope set to 10 mv/div (80 mv full scale), 5 mv offset, accuracy = {4.0% (80 mv) + 0.4% (80 mv)} = 3.52 mv horizontal range 1 ns/div to 50 s/div time base accuracy 20 ppm delay range pre-trigger: C100 % post-trigger: +100 % modes main, roll, xy xy yes reference position center trigger system sources ch 1, ch 2, ext (not applicable for tv trigger) modes normal, single, auto trigger holdoff time 60 ns selections edge, pulse width, tv edge triggers on a rising or falling edge, alternating, or either edge of any source pulse width triggers on a pulse width greater than, equal to, or less than a speci? ed time limit, with time limits ranging from 16 ns to 10 s. minimum lower limit: 8 ns minimum upper limit: 16 ns maximum pulse width setting: 10 s tv triggers on one of three standard television waveforms: ntsc, pal, secam tv trigger sensitivity: 0.6 division of sync signal. modes supported include field 1, field 2, all ? elds, or any line within a ? eld. autoscale single-button automatic setup of all channels oscilloscope channel triggering range (internal) 4 div from center screen coupling ac (< 15 hz) lf reject (~ 35 khz) hf reject (~ 35 khz) external (ext) triggering input impedance 1 m ? : _ 16 pf maximum input cat i 30 vrms, 42 vpk range dc coupling: trigger level 1.25 v and 2.5 v ext trigger pulse width > 2.5 ns
10 performance characteristics (continued) external (ext) triggering (continued) trigger level sensitivity for 1.25 v range setting: dc to 100 mhz: 100 v > 100 mhz: 200 v for 2.5 v range setting: dc to 100 mhz: 250 v > 100 mhz: 500 v display interpolation sin(x)/x display types dots and vectors persistence off, in? nite format xy, roll measurement features automatic measurements measurements are continuously updated. cursors track last selected measurement. voltage peak-to-peak, maximum, minimum, average, amplitude, top, base, vrms, overshoot, preshoot, crest, standard deviation, cycle rms, rms ac time frequency, period, +width, Cwidth, +duty cycle, Cduty cycle, rise time, fall time, delay, phase frequency maximum peak cursors modes: manual type: time, voltage and frequency (fft) measurements: ? t, ? v, frequency, peak scan (fft), ? peak math functions add, substract, multiply, fft, divide fft points 1250 points (for 500 ns and above) source of fft source channels 1 or 2 window hanning, hamming, blackman-harris, rectangular, flattop noise ? oor C50 to C90 db depending on averaging amplitude display in dbv maximum frequency 250 mhz
www.agilent.com/? nd/emailupdates get the latest information on the products and applications you select. www.agilent.com/? nd/agilentdirect quickly choose and use your test equipment solutions with con? dence. www.agilent.com/? nd/open agilent open simpli? es the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. agilent offers open connectivity for a broad range of system-ready instruments, open industry software, pc-standard i/o and global support, which are combined to more easily integrate test system development. agilent email updates agilent direc t agilent open remove all doubt our repair and calibration services will get your equipment back to you, performing like new, when prom- ised. you will get full value out of your agilent equipment through- out its lifetime. your equipment will be serviced by agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. you will always have the utmost con? dence in your measurements. agilent offers a wide range of ad- ditional expert test and measure- ment services for your equipment, including initial start-up assistance, onsite education and training, as well as design, system integration, and project management. for more information on repair and calibration services, go to: www.agilent.com/? nd/removealldoubt www.agilent.com for more information on agilent technologies products, applications or services, please contact your local agilent office. the complete list is available at: www.agilent.com/? nd/contactus americas canada (877) 894-4414 latin america 305 269 7500 united states (800) 829-4444 asia paci? c australia 1 800 629 485 china 800 810 0189 hong kong 800 938 693 india 1 800 112 929 japan 0120 (421) 345 korea 080 769 0800 malaysia 1 800 888 848 singapore 1 800 375 8100 taiwan 0800 047 866 thailand 1 800 226 008 europe & middle east austria 0820 87 44 11 belgium 32 (0) 2 404 93 40 denmark 45 70 13 15 15 finland 358 (0) 10 855 2100 france 0825 010 700* *0.125 /minute germany 01805 24 6333** **0.14 /minute ireland 1890 924 204 israel 972-3-9288-504/544 italy 39 02 92 60 8484 netherlands 31 (0) 20 547 2111 spain 34 (91) 631 3300 sweden 0200-88 22 55 switzerland 0800 80 53 53 united kingdom 44 (0) 118 9276201 other european countries: www.agilent.com/? nd/contactus revised: march 27, 2008 product speci? cations and descriptions in this document subject to change without notice. ? agilent technologies, inc. 2008 printed in usa, april 29, 2008 5989-7299en